Magnetic resonance imaging in patients with implantable devices for treatment of disturbed heart rhythm: review of the current situation

Abstract

The question of the possibility of MRI scanning in patients with cardiac implantable electronic devices (CIED) appeared simultaneously with the introduction of MRI in clinical practice. A lot of in-vitro, in-vivo and clinical researches were performed to estimate wat going on with CIED in strong magnetic field and is it possible to perform some unified protocol of safe MRI-scanning for these patients. Recommendations were provided, but not for the wide practice. MRI remained strongly contraindicated for CIED patient. To meet the clinical need CIEM manufacturers changed the design of devices to made them MRI-compatible, including reducing of ferromagnetic components, additional filters, new software. Lead coil design was changed as well to minimize lead heating and electrical current induction. Now all leaders of CIED industry have in their portfolio all types of MRI-conditional implanted cardiac rhythm management devices (pacemakers, ICDs, CRTs). “Conditional” means MRI scanning can be done only under specific condition. For MRI device and lead in one system have to be from the same manufacturer. Now, if you need to implant the device, you must proceed from the fact that the patient is more likely to need an MRI in the future and choose the appropriate model, not forgetting that the electrodes should also be MRI-compatible.

About the Author

A. A. Sviridova
Municipal Hospital #4 of the Moscow Health Care Department
Russian Federation


References

1. Бокерия Л.А., Гудкова Р.Г. Сердечно-сосудистая хирургия - 2015. Болезни и врожденные аномалии системы кровообращения. -М.: НЦССХ им. А.Н. Бакулева; 2016. 208 с.

2. Трофимова Т.Н., Карлова Н.А., Мищенко А.В., Яковлева Е.К., Халиков А.Д. Алгоритмы лучевой диагностики в системе обязательного медицинского страхования: методические рекомендации. -СПб.: Балтийский медицинский образовательный центр; 2015. 138 с.

3. Kalin R., Stanton M.S. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. Pacing Clin Electrophysiol. 2005; 28(4): 326-8. D0I: 10.1111/j.1540-8159.2005.50024.x

4. Бокерия Л.А., Гудкова Р.Г. Сердечно-сосудистая хирургия - 2014. Болезни и врожденные аномалии системы кровообращения. - М.: НЦССХ им. А.Н. Бакулева; 2015. 226 с.

5. Nazarian S., Reynolds M., Ryan M., Hunter T., Wolff S., Mollenkopf S et al. Estimating the likelihood of MRI in patients after ICD implantation: a 10-year prediction model. Journal of the American College of Cardiology. 2015; 65(10): A1090. D0I: 10.1016/S0735-1097(15)61090-7

6. Erlebacher J.A., Cahill P.T., Pannizzo F., Knowles R.J. Effect of magnetic resonance imaging on DDD pacemakers. Am J Cardiol. 1986; 57(6): 437-40.

7. Martin E.T., Sandler D.A. MRI in patients with cardiac devices. Curr Cardiol Rep. 2007; 9(1): 63-71.

8. Nazarian S., Beinart R., Halperin H.R. Magnetic resonance imaging and implantable devices. Circ Arrhythm Electrophysiol. 2013; 6(2): 419-D0I: 10.1161/CIRCEP.113.000116

9. Roguin A., Schwitter J., Vahlhaus C., Lombardi M., Brugada J., Vardas P. et al. Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace. 2008; 10(3): 336-46. D0I: 10.1093/europace/eun021

10. Sommer T., Naehle C.P., Yang A., Zeijlemaker V., Hackenbroch M., Schmiedel A. et al. Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115 examinations. Circulation. 2006; 114(12): 1285-92. DОI: 10.1161/CIRCULATIОNAHA.105.597013

11. Faris О.P., Shein M. Food and drug administration perspective: magnetic resonance imaging ofpacemaker and implantable cardioverter-defibrillator patients. Circulation. 2006; 114(12): 1232-3. DОI: 10.1161/ CIRCULATI0NAHA.106.647800

12. European Society of Cardiology (ESC), European Heart Rhythm Association (EHRA), Brignole M., Auricchio A., Baron-Esquivias G., Bordachar P. et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Europace. 2013; 15(8): 1070-118. DОI: 10.1093/europace/eut206

13. Indik J.H., Gimbel J.R., Abe H., Alkmim-Teixeira R, Birgersdotter-Green U, Clarke GD et al. 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm. 2017; 14(7): e97-153. DОI: 10.1016/j.hrthm.2017.04.025

14. Zikria J.F., Machnicki S., Rhim E., Bhatti T., Graham R.E. MRI of patients with cardiac pacemakers: a review of the medical literature. AJR Am J Roentgenol. 2011; 196(2): 390-401. DОI: 10.2214/AJR.10.4239

15. Ahmed F.Z., Morris G.M., Allen S., Khattar R., Mamas M., Zaidi A. Not all pacemakers are created equal: MRI conditional pacemaker and lead technology. J Cardiovasc Electrophysiol. 2013; 24(9): 1059-65. DОI: 10.1111/jce.12238

16. Mattei E., Triventi M., Calcagnini G., Censi F., Kainz W., Mendoza G. et al. Complexity of MRI induced heating on metallic leads: experimental measurements of 374 configurations. Biomed Eng 0nline. 2008; 7: 11. DОI: 10.1186/1475-925X-7-11

17. Ferreira A.M., Costa F., Tralhăo A., Marques H., Cardim N., Adragăo P. MRI-conditional pacemakers: current perspectives. Med Devices (Auckl). 2014; 7: 115-24. DОI: 10.2147/MDER.S44063

18. Gold M.R., Sommer T., Schwitter J., Al Fagih A., Albert T., Merkely B. et al. Full-body MRI in patients with an implantable cardioverter-de-fibrillator: primary results of a randomized study. J Am Coll Cardiol. 2015; 65(24): 2581-8. DОI: 10.1016/j.jacc.2015.04.047

19. Wilkoff B.L., Bello D., Taborsky M., Vymazal J., Kanal E., Heuer H. et al. Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment. Heart Rhythm. 2011; 8(1): 65-73. DОI: 10.1016/j.hrthm.2010.10.002

20. Mavrogeni S.I., Poulos G., Kolovou G., Theodorakis G. Magnetic resonance imaging-conditional devices: Luxury or real clinical need? Hellenic J Cardiol. 2017; DОI: 10.1016/j.hjc.2017.01.001

21. Do DH, Boyle NG. MRI in patients with implanted devices: current controversies. American College of Cardiology. 2016 [cited 2017].


Review

For citations:


Sviridova A.A. Magnetic resonance imaging in patients with implantable devices for treatment of disturbed heart rhythm: review of the current situation. Kardiologiia. 2018;58(3S):4-8. (In Russ.)

Views: 805


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 0022-9040 (Print)
ISSN 2412-5660 (Online)